Weighing Assemblies Troubleshooting

Issue Probable Cause Solution

Does not return to ZERO does not repeat known weight Bind
  • Check for platform bind.
  • Check load cell installation for plumb, square, and level.
  • Check all mounting brackets and hardware for proper torque.
Does not return to ZERO does not repeat known weight Load cell cable damaged
  • Check for broken or cut load cell cable.
  • Check for physical damage to load cell.
  • Check housing of load cell for loose or missing weigh element covers.
Erratic weights Load cell cable damaged
  • Check for broken or cut load cell cable.
  • Check for physical damage to load cell.
  • Check housing of load cell for loose or missing weigh element covers.
Erratic weights Bind
  • Check for platform bind.
  • Check load cell installation for plumb, square, and level.
  • Check load cell suspension points for free movement.
High output signal load cell Improper loading
  • Check for equal load on all load cells
High output signal load cell Load cell cable damaged
  • Check for broken or cut load cell fro cable.
  • Check for physical damage to load cell.
  • Check housing of load cell for loose or missing weigh element covers.
Low output signal load cell Load cell cable damaged
  • Check for broken or cut load cell fro cable.
  • Check for physical damage to load cell.
  • Check housing of load cell for loose or missing weigh element covers.
Low output signal load cell Improper loading
  • Check for equal load on all load cells
No output signal from load cell Load cell cable damaged
  • Check for broken or cut load cell cable.
  • Check for physical damage to load cell.
  • Check housing of load cell for loose or missing weigh element covers.
No output signal from load cell Instrument not working Properly
  • Check power supply. Check if instrument is turned on. Review Instrumentation FAQ
No output signal from load cell Bad cable connections
  • Check connections at balance box and or at instrument.
Tank / Vessel rocks from side to side Improper loading
  • Check for equal load on all load cell assemblies.
Tank / Vessel rocks from side to side Broken or damaged mounting hardware
  • Checking all mounting hardware for proper torque.
  • Check all mounting hardware for physical damage.